东芝CT机spellman电源维修CT42 CT70北京
SPELLMAN高压电源广泛应用于东芝CT机XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下
Error contents (note 1) | Cause of problem | Object part | ||
24kW | 48kW/60kW | |||
E13 | Arc- | Cathode side Arc of the X-ray tube or the high-voltage section | TUBE, INV/HV | TUBE, INV/HVC |
E25 | Arc+ | Anode side Arc of the X-ray tube or the high-voltage section | TUBE, INV/HVA | |
E12 | Over kV- | Cathode side overvoltage. Abnormal main circuit control. | INV/HV, SCB | INV/HVC, SCB |
E27 | Over kV+ | Anode side overvoltage. Abnormal main circuit control. | INV/HVA, SCB | |
E62 | Over mA- | Cathode side overcurrent of the X-ray tube or the high-voltage section. | TUBE, INV/HV, SCB | TUBE, INV/HVC,SCB |
E63 | Over mA+ | Anode side overcurrent of the X-ray tube or the high-voltage section. | TUBE, INV/HVA,SCB | |
E66 | MAC error | Filament current abnormal. Trouble of the filament heatig circuit. | INV/HV, SCB | INV/HVC, SCB |
E0 | Regulation error (HV abnormal) | There is a difference in binding output and real output | INV/HV, AC/C | INV/HVC, INV/HVA, AC/C |